Measuring The Density Of Porous Material Using Oil - OHAUS Pionnier PX85 Manuel D'instruction

Masquer les pouces Voir aussi pour Pionnier PX85:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PX Series Balance

4.5.5 Measuring the Density of Porous Material Using Oil

Press and hold the Function / Mode button until Density
appears. Press Yes to enter the Density Determination.
Item Settings:
Sample type: Solid
Auxiliary Liquid: Water
Porous Material: On
Set the following parameters by pressing No or Back:
Water Temperature
Weight
Oil Density
Measure the actual water temperature using a precision
thermometer. The balance calculates water density based on
the water temperature value entered.
Note: The weight of the sample and the density of oil must be
measured in advance.
1. Weigh the oiled sample in the air with the balance and the
density determination kit.
2. When the * (symbol of stability) appears, press the
Function / Mode button to confirm the weight of the oiled
sample in the air.
EN-21

Publicité

Chapitres

Table des Matières

Dépannage

loading

Table des Matières