Technical Data

Measuring range
Signal transmission to central controller
Supply voltage U
n
Power consumption incl. Ex-sensor
Cable gland
Cable core cross-section
Dimensions (W x H x D)
Weight
Environmental conditions
during operation:
— ambient temperature
— pressure
— humidity
during storage:
9023810 - 2nd edition - September 2005
Type XTR 0000, XTR 0001: 0 to 99 % LEL
Type XTR 0010, XTR 0011: 0 to 9.9 % LEL
Type XTR 0090, XTR 0091 with sensing head SE Ex PR M: 0 to 99 % LEL
Type XTR 0090, XTR 0091 with sensing head SE Ex LC M: 0 to 9.9 % LEL
output current 4 mA to 20 mA
12 to 30 V DC, 24 V DC nominal,
approx. 105 mA at 24 V DC
≤2.5 W
for cable diameters between 7 and 12 mm
2
0.5 to 1.5 mm
type XTR 0000, XTR 0010: approx. 80 x 150 x 60 mm
type XTR 0001, XTR 0011: approx. 110 x 150 x 60 mm
approx. 600 g (type XTR 0090, XTR 0091: approx. 450 g)
o
all types min.: –40
C
type XTR 0000, XTR 0001, XTR 0010, XTR 011
o
o
max.: T4: +65
C, T5: +55
type XTR 0090, XTR 0091: max.: T6: +65
700 to 1300 hPa
in potentially explosive atmospheres: 800 to 1100 hPa
5 to 95 % rel. humidity, non condensing
o
–40 to +65
C
700 to 1300 hPa
5 to 95 % rel. humidity, non condensing
o
C, T6: +40
C
o
C