Calibrating transmitter

For this chapter:
— Mentioning Type XTR 0000 also includes type XTR 0001 as well as the transmit-
ters type XTR 0090 or XTR 0091 with remote sensing head SE Ex PR M
(100 % LEL).
— Mentioning Type XTR 0010 also includes type XTR 0011 as well as the transmit-
ters type XTR 0090 or XTR 0091 with remote sensing head SE Ex LC M
(10 % LEL).

Zero point adjustment

To enter the calibration menu press the key
than 3 seconds. The display shows a flashing point on the right bottom indicating the
calibration mode.
Navigate by means of keys
the menu by pressing the keys (
the calibration procedure mode to the controller unit.
Apply zero gas to the sensor –
without calibration adapter:
Ensure that the transmitter is in clean ambient surroundings (no flammable gases
or vapours present),
or
with calibration adapter:
Supply clean air at a flow rate of approx. 0.5 L/min via the calibration adapter.
— The transmitter's display shows the currently measured gas concentration which
in general is slightly different from zero. Negative values down to "–9" % LEL or
"–.9" (type XTR 0010) are displayed.
Wait until the reading has stabilised (max. 3 minutes) and press keys (
this the current sensor signal is converted into an up-dated zero point also shown
on the display (ideally "0").
q
l
Press keys (
+
) to save the up-dated zero point and to terminate the zero adju-
stment function.
Remark:
After a sensor replacement the zero point might be misaligned so much
that the display shows "--". In this case press keys (
nary zero and again follow the above zero adjustment procedure.
q
l
Press keys (
+
) to exit this function. The display shows ZE again.
q
Press key
to switch back to the measurement mode or press key
the sensitivity adjustment.
9023810 - 2nd edition - September 2005
l
for more than one second but less
q
l
or
until the display shows ZE ("ZEro") and enter
q
l
+
). The output signal is set to 3.4 mA to signal
q
l
+
). By
q
l
+
) to save prelimi-
l
to perform
Calibrating transmitter
Page 17 of 42