15.1
Warnings

15.1.1 General information

NOTICE
Do not clean the product in the ultrasonic bath.
Otherwise, the product could be damaged.
Do not clean the product with a metal brush or steel wool.
Otherwise, the product could be damaged.
15.1.2 Cleaning and Disinfection Agents
Do not use cleaning and disinfection agents that contain the following substances:
• Caustic soda
Do not use any rinse agents.
For rinsing only use water with the following properties:
• Low in endotoxins (max. 0.25 endotoxin units/ml; e.g., purified water).
• Sterile / low in germs (max. 10 germs/ml).
Only use agents suitable for cleaning / desinfection of plastic and metals.
15.2
Limitations on Reprocessing
Frequent reprocessing has little impact on these instruments. The end of the
product lifetime is usually based on wear and tear as well as damage from use.
15.3
Pre-treatment at the Point of Use
Clean the product immediately after use from rough impurities, corrosive
solutions and drugs. To do this, rinse the product in flowing cold water and wipe it
clean.
15.4
Preparation for Cleaning
Start:
Preparation:
Cleaning solution:
Soaking in cleaning solu-
tion:
Rinse:
20
As soon as possible, but at the latest 16 h after
product use
Open the product as widely as possible / disas-
semble its individual parts.
Cideyzyme (Johnson & Johnson), 0,5%, 20 °C
Duration of soaking: 5 min
To start the soaking time rinse all lumina at least
twice using a single-use syringe (min 5 ml) with
cleaning solution. Brush all surfaces using a suitable
brush.
Brush all lumina using a bottle brush in flowing wa-
ter.