Configuring

Baud
4 0
8
pH
C
A L
i
p
d s
p
9
4
pH
1
n
T 3
Time
d
O
1
n
T 4
Time
d
9
0 0
Day.Month
12
Configuring
Note
You can leave the configuration menu at any time with <M>. The parameters
already changed are stored.
Configuring procedure (factory settings appear in bold typeface):
l Switch off the measuring instrument.
l While pressing the <M> key, press <
l Select the required Baud rate with < > < >.
0
Selection: 1200, 2400, 4800, 9600 Baud.
l Then press the <RUN/ENTER> key.
l Select the required display during the pH calibration with < > < >.
/pH
Selection: Buffer nominal value (/pH) or current electrode voltage (mV).
l Then press the <RUN/ENTER> key.
l The air pressure is only displayed if an oxygen sensor is connected. Oth-
erwise, "---" appears.
mbar
l Continue with the <RUN/ENTER> key.
9
l Select the required calibration interval for pH/ORP measurements with
< > < >.
Selection: 1... 7... 999 d.
7
l Then press the <RUN/ENTER> key.
l Select the required calibration interval for oxygen measurements with
< > < >.
Selection: 1... 14... 999 d.
1
4
l Then press the <RUN/ENTER> key.
l Select the date and time step-by-step with < > < >. In between, press
4
the <RUN/ENTER> key each time.
l When the last <RUN/ENTER> key has been pressed, the instrument
changes to the last active measuring mode.
>.
pH/Oxi 340i