5 - ELECTRICAL SPECIFICATIONS

Reference conditions
Temperature: +20°C to +26°C
Relative humidity: 45% to 75% RH
Battery voltage: 8.5 V to 9 V
Position of the conductor: centred in the clamp
Absence of external AC magnetic field
Absence of electric field
Measurement on sinusoidal signal from 45 Hz to 65 Hz
Measurement extent on AC
From 45 Hz to 450 Hz
Accuracy
The intrinsic error is expressed as a ± % of reading (rdg) and in measurement counts
(counts).
5-1 A AC RMS AND A DC CURRENT
Measurement extent
Range
A AC RMS
(1)
400 A (2)
0.2 A...399.9 A
1000 A (3)
400 A...1000 A
(1) On automatic and manual
(2) From 0.2 A to 50 A: 2% rdg ± 10 counts
(3) Accuracy given from 0 to 90% of the range
From 90 to 100% of the range: 5% max
NB : It is the form of the jaws, thus the cable clamping capacity, that limits the
current range of the clamp. The measurement range is greater.
(4) Supplementary error with crest factor in A AC RMS (< 1500 A peak)
Fc from 1.5 to 3: 5% rdg
Fc from 3 to 5: 12% rdg
(5) In RMS A AC up 800 A to 1000 A, accuracy 5 % rdg ± 5 pt
In A DC up 1000 A to 1400 A, accuracy 8% rdg ± 5 pt
(6) Supplementary error in A AC, depending of frequency (from 65Hz to 450Hz) : < 5%
Permitted overload:
Resolution
A DC
± 399.9 A
100 mA
± 1400 A
1 A
1000 A RMS permanent up to 1 kHz
Temperature of conductor ≤ 90°C
36
Accuracy
(4) (6)
2.5% rdg
±2counts
(5)