→ Press Mode key 4 for scrolling more display function as follows:
Press emissivity key 4 key to adjust the emissivity. The emissivity
can be changed from 0.10 to 1.00.
Press the Meas. key 6 to display the surface temperature.
Keep pressing Mode key 4 for the Maximum (MAX), Minimum
(MIN), Different between MAX and MIN (DIF) and Average (AVG)
readings during the last measurement period.
Press Down key 2 or Up key 3 to change the High Alarm (HAL) or
Lo Alarm (LAL). F. e.: When the reading 27 °C < LAL 27.1 °C, the
Low icon will flash and you will hear a beep sound.
The measurement range: -60 .. +500 °C.
→ ScanTemp will automatically shut off if left idle for more than 60 sec,
unless in PRB mode.
Thermocouple probe scan function:
Press the mode key 4 until PRB appears: on the LCD.
Connect the thermocouple with Scan Temp 450's thermocouple socket 5 and put
the probe in or on the target, the temperature will be displayed beside the "PRB"
icon.
→ ScanTemp 450 will automatically shut off if left idle for more than 12 min in
PRB mode.
→ The measurement range: -64 to +1400 °C.
After measure high temp, the probe may remain HOT for a while.
12
GB
Key pad:
In Default:
Press Lock key for infrared LOCK mode ON/OFF.
Max, Min, DIF, AVG,
Press °C/°F key for °C or °F transferred.
HAL, LAL, PRB
In all modes: Hold the
Press Lock key for backlight function ON/OFF.
Measuring key 6
Press °C/°F key for Laser function ON/OFF.
Batterie:
Battery OK
Battery Low
Measurements are
Battery needs to be
possible
replaced, measurements
are still possible
When the 'Low Battery' icon indicates the battery is low, the battery should be
replaced immediately with two AAA, 1.5V batteries.
Please note: It is important to turn the instrument off before replacing the battery
otherwise the ScanTemp 450 may malfunction..
Battery Exhausted
Measurements are not
possible