Chapter 18
TROUBLESHOOTING
Abnormal Phenomenons
BP measurement values
too high or too low
No pressure
Cuff deflates in short time
It can not carry on measurement when press the measurement
button
Power off suddenly when
inflating
Causes
Cuff is not connected correctly
Talk or move arms when
measuring
The turnup clothing presses
the arm
Cuff leakage
The cuff windpipe is not correctly
connected with cuff
Cuff is not inflated
Loose cuff
No use for a long time,
the power of batteries
can be exhausted owing
to the changed temperature
ENGLISH
Solutions
Correctly connect cuff
(refer to Chapter 9)
Keep quiet and restart
a measurement
Take off the clothing which
presses the arm, and restart
a measurement
Buy a new cuff
Correctly connect
Stop using the device
and contact us
Correctly apply cuff
Switch on the power once again
and restart a measurement
Replace all four batteries
with new ones