Tableau 19 Paramètres de bus du bloc de transducteurs (suite)
Parameter
PRIMARY_VALUE_TYPE
FILTER_TIME
IMPEDANCE
SENSOR_CURRENT
Temperature
SECONDARY_VALUE_2
SECONDARY_VALUE_
UNIT_2
Description
Selects the displayed primary
value depending
on the selected measuring
method in APPLICATION
parameter
Sets the filter time of the input
filter
Shows the impedance of the
sensor
Shows the momentary current
of the sensor
Process temperature value and
status
Value
Status
Degree C or degree F. Changes
the unit of temperature being
displayed and transmitted.
Data
R/W Bytes
Range
type
0 = DO_%
1 = DO_mg/l
2 = DO_g/l
R/W 2
uns16
3 = DO_ppm
4 = DO_ppb
5 = GAS_ppm
6 = GAS_%
0–120
R/W 2
uns16
Default: 0s
R
4
float
R
4
float
R
4
DS_65
R
1
1001 = °C
R/W 2
uns16
1002 = °F