Hach si792 D Manuel D'utilisation page 64

Masquer les pouces Voir aussi pour si792 D:
Table des Matières

Publicité

Tableau 19 Paramètres de bus du bloc de transducteurs (suite)
Parameter
COLLECTION_
DIRECTORY
Output
INPUT_RANGE
POLARIZATION_VOLTAGE Sets the polarization voltage
SENSOR_TYPE
APPLICATION
PRIMARY_VALUE
Description
A directory that specifies the
number, starting indices, and
DD item of IDs of the data
collection in each transducer
within a transducer block. Used
by the host for efficient transfer
of information.
Selects the connection of the
sensor
Selects the used sensor type.
Selects the measurement
method
(DO/ GAS)
Shows the primary value and
status
Value
Status
Data
R/W Bytes
Range
type
R
36
0 = LO_LEVEL
R/W 2
uns16
1 = HI_LEVEL
R/W 4
float
–1000–0
0 = Standard
R/W 2
uns16
1 = Traces
0 = Dissolved
oxygen
R/W 2
uns16
1 = Measurement in
gases
4
R
DS_65
1