Move data to a USB memory stick
1. Connect the memory stick to the USB port on the bottom of the instrument.
2. Navigate to the Buffered Data Icon and push the Select button.
3. In the Buffered Data screen, push the Left arrow until the icon that shows the USB stick and a
green arrow is highlighted. Push the Select button.
The data export starts automatically. The display shows a successful export message when the
export is complete. Do not remove the memory stick before this message shows. For buffers with
a large amount of data, the export may take a few seconds.
4. Push the Select button to remove the message.
5. Remove the USB stick from the instrument and connect the stick to the USB port on the
computer.
6. In Windows Explorer, navigate to the computer drive for the USB memory stick.
7. Right click the DATA.tsv file (or the secure .xls file if enabled in the Communications Setup
screen) and select Open with > Excel. When the file is open, move the data to the computer.
Note: Each time data are saved to the USB memory stick, the DATA file on the memory stick is overwritten.
Save data with an Ethernet connection
Note: This option is available only when the instrument is used with a charge cradle. The cradle is an optional item
for some instrument models
1. Connect the Ethernet cable and power cable to the instrument base.
2. Navigate to the Communication Setup menu and push the Select button.
3. In the Communication Setup screen, set the IP address, Subnet Mask, and Gateway address. To
do this, do one of the following:
a. Put the IP Address, Subnet and Gateway data in the appropriate fields OR
b. Select the DHCP check box.
If the DHCP box is selected, the IP Address, Subnet Mask and Gateway are set automatically
when the instrument is connected to the network. The instrument will ask for an IP address if:
• the instrument is put in an Ethernet connected cradle.
• the instrument is in a cradle and a power cycle is done.
• the user exits from the communications screen after changes were made but the DHCP
box was still checked.
4. Install the instrument in the cradle.
5. Open the Internet browser. In the Address bar, put one of the items that follow:
a. The IP address from the Communication Setup screen
b. hpc+ the instrument serial number. The serial number is in the instrument Diagnostic screen.
Example: hpc123456789
The HPC Data Web Server page opens.
Note: For this option, the PC and the HHPC+ instrument must be connected to the same local area
network. The PC and the HHPC+ instrument must not be separated by a router, and the network must be
configured to allow packet broadcasting. In case of difficulty, contact your network administrator.
c. Press the "rebuild button" on the browser page to build the file.
6. Open the DATA.tsv file or if enabled the secure DATA .XLS file.
Move data to a mass storage device
1. Connect the mini-USB cable to the mini-USB port on the instrument and the USB port on the PC.
2. In Windows Explorer on the PC, navigate to the HPC USB drive and open the DATA.TSV file.
3. Disconnect the USB cable or do a power cycle of the instrument.
New data is added to the DATA file.
16 English
(HHPC models
on page 6).