Vercise™ DBS Systems Remote Control 4 Handbook, 92753307-07
Stimulator Error Screen
If the system check fails due to a Stimulator
error, MRI Mode will not be enabled and the
Remote Control displays the Stimulator Error
screen. This screen may display when there is
a communication error between the Remote
Control and Stimulator.
Page 48
Press
to acknowledge the message and
retry entering MRI Mode. If this error screen
immediately displays again, your healthcare
provider should not perform an MRI scan.
Contact the healthcare provider managing
your DBS System or Boston Scientific.