4. Safety instructions:
SIM/SD-Card replacement is only allowed in ordinary (non-hazardous) locations.
■
The device may only be charged in a temperature range of -5°C to 55 °C.
■
Charging and wired data-transfer via the USB-C interface is limited to a maximum
■
Um of 30V and a current of 50A (Im).
It must be ensured that the power plug/supply used for charging is a non-shock hazard
■
extra low voltage equipment e.g. SELV, PELV or ES1 as per IEC 62368-1
(former IEC 60950-1) or equivalent.
Intrinsically safe accessory certified for use in hazardous locations must match with the
■
entity parameter of the USB interface.
Entity parameter USB-C interface for use in hazardous locations
Uo = 5.6 V / Io = 500 mA / Po = 600 mW / Co = 1 μF / Lo = 1 μH
Ui = 5.6 V / Ii = 100 mA / Pi = 100 mW / Ci = 53 uF / Li = 0 μH
Spare parts user replacement in ordinary locations
The relevant spare part replacement instructions are provided on the manufacturer's
homepage.
The following spare parts are allowed to be replaced by the end user.
USB-C Cover Tab-Ex 03 DZ1 – (compatible with Tab-Ex 05 DZ1)
■
5. Repair
Repairs are to be conducted by the manufacturer or an authorized service centre.
6. Optional non electrical accessories
The following non-electrical accessory can be used with the Tab-Ex 05 DZ1 in the hazardous
classified location, provided that all specified restrictions and conditions are observed.
S Pen type ST T05 X2
■
Leather case type LC T05 X1
■
Chest Harness type CH T05 X1
■
Belt clip type BC T01
■
Carrying strap CS T05
■
Tether for stylus pen type ST 01
■
4