Télécharger Imprimer la page

ResMed AirCurve 11 VPAP Tx Mode D'emploi page 12

Publicité

Using the touch screen:
There are two actions to navigate through the touch screen:
Swipe: Swipe up or down the screen to view menu options.
Tap: Select a parameter setting to update. For other parameters (eg Pressure Relief, Airplane mode), tap
the parameter to turn it on
Personalizing settings
After the initial set-up, clinicians have the ability to personalize settings to make therapy more comfortable.
These settings are updated through the MY OPTIONS menu.
1. Tap MY OPTIONS from the Home screen.
2. Tap the parameter to update.
3. Tap the preferred setting.
4. Tap OK to confirm the change or CANCEL to go back to the previous screen.
Additional features
There are some other features on the device which the clinician can personalize.
Note: Not all functions ar e availa ble in all regions. Functions vary based on therapy m ode.
Menu
Function
Ramp Time
MY
OPTIONS
Pressure
Relief*
Mask
Tube
SmartStart™* When SmartStart is enabled, therapy starts automatically when the patient
SmartStop*
MORE
Warmup
Mask Fit
Device
Diagnostic
* Feature must be enabled by the care provider before the patient can use.
or tap to turn it off
Description
Period during which the pressure increases from a low start pressure to the
prescribed treatment pressure.
Ramp Time can be set to Off or 5 to 45 minutes (in 5-minute increments).
When EPR (Expiratory Pressure Relief) is enabled, the patient may find it
easier to breathe out. This setting can help the patient get used to therapy.
Select the type of mask used with the device.
Select the type of tubing used with the device.
breathes into the mask.
When SmartStop is enabled, therapy stops automatically after a few
seconds when the patient removes their mask.
This function will heat the water in the humidifier tub
This function assesses and identifies possible air leaks around the mask.
Indicates good mask seal. Leak is less than 24L/min.
Adjust the mask. Leak is more than 24L/min.
When enabled, Device Diagnostics will analyze the functionality of the
device. See ResMed.com for further details.
Device Diagnostic can be set to run Daily, Weekly, every 2 weeks, or it can
be switched off.
.
English
9

Publicité

loading