Measure
Initio 2 conductivity meter
Ref : 701 653
2.1 Start up
The measurement is displayed directly on the instrument screen.
2.2 Browsing
Initio 2 has three buttons for browsing :
Selection-validation (3):
The central knob (3) is used to browse through the Initio 2 menus. Do a brief press on the
OK button on the middle of the knob to validate.
Return (2):
The "Return" button (2) changes from one type of display to the other from the main screen
in the following order:
Within a particular menu, it acts as a Back button.
Menu (4):
This button displays the Initio 2 main menu.
2.3 Range
The Initio 2 Conductivity meter has two ranges
To change from one to another:
2.4 Use with the CTM probe
The CTM probe does not have a built-in temperature sensor. However, the measurement
can be offset making use of an external thermometer when the calibration is made:
With external thermometer
Without external thermometer
ENGLISH
-
Press the power on button (1) on the left side of the instrument.
-
To switch the Initio 2 off, do a short press on the power on button (1) and validate
pressing the knob (3).
-
Measurement
-
Table
-
Graph
-
2 mS/cm
-
20 mS/cm
-
Go to the main screen (measurement display) using the Return button
-
Press the validation knob
-
Select the required range
-
Validate by pressing the knob
Press the Menu "button"
-
-
Select Calibration
-
Enter the temperature read with the thermometer using the rotary knob
-
Select Calibrate
-
Enter the value of the standard solution at 25°C and validate
Press the Menu "button"
-
-
Select Calibrate
-
Select Without offsetting