Procedure for
10
disinfecting the sensor
Pre-disinfection
cleaning
Immersion in an enzymatic
cleaning solution
(Aniosyme DD1, Hexanios G
+ R or surface cleaning
((Anios disposable cleaning
cloth)).
Any other disinfection method is prohibited. Damage caused by these other methods cannot be
covered by the manufacturer.
Use of a sterile protective cover on the video camera is recommended regardless of the process
adopted by the sterilisation service, with this applying throughout the surgical operation.
The video camera cannot be placed in an autoclave.
The video camera is not compatible with automatic washing and disinfecting machines.
Existing alkaline solutions for the preliminary disinfection of certain medical devices are PROHIBITED
for the purpose of preliminary disinfection of our video cameras.
You absolutely must rinse copiously those parts that have come into contact with the disinfectant.
In order to avoid scratching them, use non-woven compresses for drying the lenses.
The procedures described in this section are provided as advice; under no circumstances may they
take the place of offi cial directives or recommendations.
As decontamination depends on the products, methods and/or tools selected, this is entirely the
responsibility of the staff in question.
011546C
Disinfection
Immersion in a glutaraldehyde
solution (Anios Laboratory -
Steranios 2%)
Serilisation
• Sterilisation using the Sterrad
100NX process standard cycle (47
minutes)
• Sterilisation usign ethyle oxide
19