Installation of the video
camera
This medical device is intended for use by a qualifi ed surgeon as part of endoscopic explorations.
Installing it does not require any special training. Please refer to the instructions in this manual. Place the
communication channel on a stable support and take its various accessories.
• Connect the power supply cord to the video camera's mains socket [C2].
• Connect a video cable to one of the video outputs: Preferably, use the Y/C output [C8], the SDI output
[C3 & C4], or the RGB output [C7], which provide a better result than the composite output [C9].
• Connect the second end of the video cable to the monitor's video input.
Connect the video camera's USB2 output [C5] to the computer's USB2 input.
• Place the switch [S9] in position 1. This will place the video camera in stand-by mode (the green
indicator [L1] should then fl ash).
• Hook up the sensor connector to the front panel of the control channel unit [C3].
• Power up the monitor.
• Press the button [S1] to start up the video camera (the green indicator will light up continuously).
Pressing it once more will return it to stand-by position.
• An image will appear on the monitor and a driver installation message will appear on the computer;
when it does, select the driver delivered with the video camera.
It is prohibited to carry out a surgical operation with the USB2 output; a video output must be
connected to the monitor.
Only the EMPIA2820 driver should be installed for the video camera's USB2 output. (The driver is
available with the CD delivered.)
Then it will be possible to check that the video camera is operating properly and in particular to make
sure that the monitor's settings are completely satisfactory: when you disconnect the sensor, a bar test
pattern will appear (except on the USB outlet [C5]).
Before fi tting an endoscope (or a fi brescope) to the front of the camera lens, you should check the
condition of the endoscope in question (check that the distal lens or eye-piece is clean).
The functions can be accessed using a single adjustment wheel [S2]. Turning the wheel allows you to
browse through the Menu and pressing on the wheel enables you to validate an item.
12
011546C