pH Configuration
Prior to measurement, a configuration should be performed
to match the connected sensor and the desired measurement
performance. Furthermore, you can select the suitable calibra-
tion method.
The following table gives you an overview. Factory settings are
shown in bold print.
Measurement
"Setup" display
Display 1
Display 2
CAL Timer
CAL
CAL POINTS
BUFFER SET
(AutoCal,
FREE CAL)
Auto OFF
Temp Unit
Time Format
Date Format
Default
This icon prompts you to select a menu item using the arrow keys –
the selection is confirmed by pressing set.
Select using arrow keys, confirm by pressing set.
pH x.xx / pH x.xxx / mV /
OFF / date + time / date / time
OFF / 1 ... 99 days
AutoCal/Manual/DATA INPUT/(ISFET-Zero)/FREE CAL
1 / 2 / 3 / 1-2-3 (for AutoCal, Manual, FREE CAL)
-01-
(De-
NIST standard
fault)
-02-
NIST technical
SI Analytics
-03-
technical
-04-
Knick CaliMat
-05-
Mettler Toledo
-06-
HACH
-07-
Ciba (94)
-08-
Hamilton
-09-
Reagecon
-10-
DIN 19267
-U1-
loadable via HandyLab Pilot (User)
OFF/ 0.1h / 1h / 6h / 12h
°C / °F
24h / 12h
dd.mm.yy / mm.dd.yy
NO / YES (reset to factory settings)
Note: All data logger entries will be deleted.
Configuring
(°C for analog pH only)
1.679/4.006/6.865/9.180
1.68/4.00/7.00/10.01/12.46
2.00/4.00/7.00/10.00
2.00/4.00/7.00/9.00/12.00
2.00/4.01/7.00/9.21
4.01/7.00/10.01/12.00
2.06/4.00/7.00/10.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
41