3 Getting Started
Set the Output Voltage
Method 1
Use the left and right navigation keys to navigate to the setting that you wish to change.
In the following display, the voltage setting is selected. Enter a value using the numeric keypad. Then
press Select.
You can also use the numeric arrow keys to adjust the value up or down. Values become effective when
the output is turned on.
In voltage priority mode, the unit will maintain the output voltage at its programmed setting. In current
priority mode, the unit will limit the output voltage when it reaches the specified voltage limit value.
Refer to
Set the Output Mode
for more information.
Method 2
Use the Voltage key to select the voltage entry field. In the display below, the voltage setting is
selected. Enter the desired setting using the numeric keypad. Then press Enter.
If you make a mistake, either use the backspace key to delete the number, press Back to back out of
the menu, or press Meter to return to meter mode.
Keysight RP7900 Series Operating and Service Guide
99