Les langues disponibles

5 SCPI Programming Reference
Data was sent in arbitrary block format but is not allowed for this command.
Execution Errors (these errors set Standard Event Status register bit #4)
-200 Execution error
Generic syntax error
-220 Parameter error
A data element related error occurred.
-221 Settings conflict
A data element could not be executed because of the present instrument state.
-222 Data out of range
A data element could not be executed because the value was outside the valid range.
-223 Too much data
A data element was received that contains more data than the instrument can handle.
-224 Illegal parameter value
An exact value was expected but not received.
-225 Out of memory
The device has insufficient memory to perform the requested operation.
-226 Lists not same length
One or more lists are not the same length.
-230 Data corrupt or stale
Possible invalid data. A new reading was started but not completed.
-231 Data questionable
The measurement accuracy is suspect.
-232 Invalid format
The data format or structure is inappropriate.
-233 Invalid version
The version of the data format is incorrect to the instrument.
-240 Hardware error
The command could not be executed because of a hardware problem with the instrument.
-241 Hardware missing
270
Keysight RP7900 Series Operating and Service Guide