5 SCPI Programming Reference
Parameter
BUS|CURRent1 | EXTernal | PIN<1-7> | TRANsient1
| VOLTage1, *RST BUS
Select digital port pin 1 as the measurement trigger source: TRIG:ACQ:SOUR PIN1
TRIGger:ARB:SOURce <source>
TRIGger:ARB:SOURce?
TRIGger:ELOG:SOURce <source>
TRIGger:ARB:SOURce?
TRIGger:TRANsient:SOURce <source>
TRIGger:TRANsient:SOURce?
TRIG:ARB:SOURce - Selects the trigger source for arbitrary waveforms:
TRIG:ELOG:SOURce - Selects the trigger source for external data logging:
TRIG:TRANsient:SOURce - Selects the trigger source for the transient system:
BUS
Selects a remote interface trigger command.
EXTernal
Selects ALL digital port pins that have been configured as trigger sources.
IMMediate
Triggers the transient as soon as it is INITiated.
PIN<1-7>
Selects a digital port pin configured as a trigger input.
Parameter
BUS | EXTernal | IMMediate | PIN<1-7> *RST BUS
Select digital port pin 1 as the Arb trigger source: TRIG:ARB:SOUR PIN1
Select digital port pin 1 as the transient trigger source: TRIG:TRAN:SOUR PIN1
246
Typical Return
BUS, CURR1, EXT, PIN<n>, TRAN1, or VOLT1
Typical Return
BUS, EXT, IMM, PIN<n>