EGAmaster EGATRONIK Manuel D'instructions page 30

Publicité

Les langues disponibles

Les langues disponibles

2. Set the rotary switch to ºC.
3. Place the temperature probe to the object being measured.
The measured value shows on the display.
Note
- The Meter automatically displays the temperature value inside the Meter when there is no
temperature probe connection.
- The included temperature probe can only be measured up to 250. For any mesaurement higher
than that, the rod type temperature probe must be used instead.
- When temperature measeuremnet has been completed, disconnect the connection between the
testing leads and te circuit under test.
Warning!
To avoid damage to the Meter or to the equipment under test, disconnect circuit power and
discharge all high-voltage capacitors before measuring capacitance.
Use the DC voltage function to confirm that the capacitor is discharged. Never attempt to
input over 60V in DC or 3V rms in AC to avoid personal dangerous.
Capacitance measurement has 4 measurement positions on the rotary switch : 2nF, 20nF, 2µF
and 20µF.
To measure capacitance, connect the Meter as follows:
1. Insert the red test clip or red test lead into the the
or black test lead into the mA terminal.
2. Set the rotary switch to an appropriate measurement position in Fcx range.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
30
All manuals and user guides at all-guides.com
(figure 8)
I. CAPACITANCE MEASUREMENT (see figure 9)
ºC terminal and the black test clip