OBJ_DOKU-169544-013.fm Seite 41 Mittwoch, 17. März 2021 4:45 16
• Cleaning/disinfection by machine in a STERIS Reliance Gen-
fore washer/disinfector GEN FH07-1XX (STERIS Corporation,
Mentor OH, USA) using neodisher mediclean forte detergent
(Dr. Weigert GmbH & Co. KG, Hamburg/Germany).
– Prewash: 2 minutes under cold tap water.
– Main wash: 5 minutes at 140°F (60°C) with neodisher
MediClean forte, dosage ¼ oz/gallon.
– Final rinse: 1 minute under hot tap water.
– Product drying: 7 minutes at 194°F (90°C).
– Disinfection: 5 minutes at 194 °F (90 °C).
• Sterilization in an autoclave Systec V-150 (Systec GmbH,
Linden/Germany).
– Steam sterilization with saturated steam, fractionated
vacuum procedure, 3 minutes, 270 °F (132 °C), 44 psi
(3 bar).
Erbe recommends the processing procedures described in these
instructions for use. Equivalent different procedures are possible
if not explicitly excluded. It is incumbent on the user to ensure
the suitability of the actual procedures used by means of suit-
able measures (e.g. validation, routine monitoring, check of ma-
terial compatibility).
6
Disposal
Dispose of the product, packaging material and accessories (if
available) in accordance with the national guidelines and regu-
lation applicable in each case.