5
Cleaning, Disinfection, Sterilization
5.1
Safety instructions for processing
May be cleaned in an ultrasonic bath.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 29 psi (2 bar).
Maximum water pressure during rinsing: 29 psi (2 bar).
During machine cleaning and disinfection, do not exceed 203 °F
(95 °C).
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
During sterilization, do not exceed 280 °F (138 °C).
Do not sterilize in hot air oven.
5.2
Processing Limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
5.3
Disassembly
1. First, separate the cable from the instrument. Do not damage
the cable insulation while doing so.