Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
1. Roll up the product loosely and lay it in a suitable rinsing
basket. In doing so make sure the product is not touching
other instruments or parts of instruments.
2. Start a tested program with the following properties:
– A cleaning cycle consisting of prewash, main wash, final
rinse and product drying.
– Final rinse with hot tap water (at least 1 minute).
– Sufficient product drying (no visible fluid residues).
– Thermal disinfection (5 to 10 minutes at 194 to 203 °F
[90 to 95 °C], A
≥ 3000).
0
3. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
5.8
Check
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
If damaged, do not use this product!
5.9
Packaging
1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/polyethylene and/
or in a sterilization container.
39