OBJ_DOKU-169544-013.fm Seite 22 Mittwoch, 17. März 2021 4:45 16
1.5
Performance characteristics
Performance characteristics relating to the intended use are:
• Transporting HF current from the electrosurgical unit to the
instrument.
2
Maximum electrical capacity
The maximum electrical capacity of these products is:
• Number Group 20192-XXX: 5000 V
• Number Group 20196-XXX: 1000 V
3
Safety instructions
WARNING!
Check insulation for damage before every application.
Erbe recommends carrying out an electrical continuity test be-
fore each application of these products.
If the cable conductor is broken or the electrical transmission of
the cable is interrupted in any other way, this may lead to arc
formation in the patient return line or in the active circuit; this
may cause burns to the patient/surgical staff or start a fire.
If damaged, do not use these products!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Damp inside the connectors may cause flash-overs that could
damage the connector or its insulation. Ensure that the product
is dried sufficiently during processing and prevent fluids from
penetrating the connector during use. The older the material, the
greater the risk of unwanted damp in the connectors.
p
p
22