During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
5.2
Processing limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
5.3
Dismantling
1. First, separate the cable from the instrument. Do not damage
the cable insulation while doing so.
5.4
Required aids
Processing step
Items Needed
Precleaning
Soft plastic brush/soft cloth
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
Compressed air gun (2 bar maximum)