Problem
Reason
Readings from
Blood pressure is a
repeated mea-
fluctuating value.
surements differ
For healthy adults,
considerably.
deviations of 10 to
20 mmHg are possible.
Cuff is not properly fitted.
Readings were not con-
sistently taken at heart
level.
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
Prior to a mea-
Position sensor is de-
surement, the
activated.
position sensor
does not work.
All manuals and user guides at all-guides.com
Solution
none
Ensure that the appliance is fitted
as follows:
• on top of the slim side of the
forearm
• with the display on top
• with the battery compartment
facing the user's body
• approx. 2 cm away from the thumb
root
For each measurement, proceed
exactly as shown and described
under «Measuring».
While measuring, relax, keep still,
do not move, or speak.
Refer to the chapter «Posotion
Sensor» to activate the position
sensor.