Command: READ PUMP STATUS
Source: PC
Destination: Pump (with address = 3)
02
STX
Source: Pump (with address = 3 in stop status)
Destination: PC
02
STX

Window Meanings

The following table describes the windows available to the user.
The columns R/W indicates if the window is read-only (R) or if it can
also be modified (RW).
The windows default values are shown in bold.
WIN
TYPE
R/W
000
L
RW
008
L
RW
102
N
RW
105
N
RW
108
N
RW
120
N
RW
201
N
R
202
N
R
203
N
R
205
N
R
MS120 Single Stage Rotary Vane Pumps User Manual/87-900-150-01 (C.00)
83
32
30
35
ADDR
WINDOW
83
32
30
35
ADDR
WINDOW
DESCRIPTION
START/STOP (1= START; 0= STOP)
REMOTE/SERIAL CONFIGURATION (1= Remote; 0= Serial)
REMOTE I/O SET POINT VALUE [Hz] (0 – 127) 40
REMOTE I/O SET POINT HYSTERESIS [%] (0 – 100) 2
SERIAL COMMUNICATION BAUD RATE (0-6)
[0=600, 1=1200, 2=2400, 3=4200, 4=9600, 5=19200 bit/s]
ROTATIONAL FREQUENCY SETTING [Hz] (40 - 60) 58
3PHASE VOLTAGE [V] 3phase Voltage supplied to the motor
POWER [W] Electrical power supplied to the motor
DRIVING FREQUENCY [Hz] Reads the actual electrical driving frequency
STATUS (0=Stop; 1=Wait interlock; 2=Start; 3=Autotuning; 5=Normal;
6=Fail)
Technical Information
30
03
38
37
RD
ETX
CRC
30
30
30
30
30
DATA (STATUS)
5
User Interface
30
03
38
37
ETX
CRC
185/193