The pump must not be installed outdoors and must always be protected from the
CAUTION!
weather.
Avoid warm air coming from the exhaust or the cooling fans causing discomfort
to the personnel.
Do not install the pump in a dusty area or where other materials may block or
cover the cooling surfaces quickly.

Connection to the machine

The connection to the chamber to be pumped down must be carried
out by means of pipes of the same diameter as the inlet port.
MS120 Single Stage Rotary Vane Pumps User Manual/87-900-150-01 (C.00)
Minimum distance
A
150 mm
Figure 9
Instructions for Use
Preparation for Installation
B
C
200 mm
150 mm
4
D
500 mm