Screwing the oil drain pipe on the discharge (G), the lubricant will
start to drain. It is recommended to tilt the pump forwards by raising
the back 40 mm to ensure the complete draining of the lubricant,
remember to brake the wheels to prevent the pump from falling. Once
the oil in tank has completely been discharged from the tank,
reassemble both plugs ("E" and "G") and let the pump run under
vacuum for about one minute, so that the lubricating line gets
emptied and any oil residual keeps inside the pump. Then remove the
plugs and discharge the rest of the oil.
MS120 Single Stage Rotary Vane Pumps User Manual/87-900-150-01 (C.00)
Figure 14
Figure 15
Instructions for Use
Maintenance
153/193