Caution Notes

a
Before putting the pump back into operation after a breakdown, inspect it and check
carefully for any other signs of damage.
b
Use only tools that are in perfect working order and specially designed for the job; use of
inappropriate or ineffective tools may cause serious damage.
c
Always check the lubricant and that it is properly distributed through the pump;
inadequate lubrication may damage the pump seriously.
d
Give the parts some form of marking as you strip them down to ensure that you
reassemble them again in the proper order.
e
Check that there are no scratches or grooves on the machined shafts, in their seats
inside the pump or on machine-ground surfaces. Slight scratches and abrasions may be
eliminated with very fine emery paper or by a little light grinding.
f
Before putting a group together, always spread a little oil over inner parts and mating
surfaces. Replace all seals with original spare parts before reassembling components.