TP5000HY
1-3 Electrical Specifications
Accuracy is given as ± (% of reading + counts of least significant digit) at 23°C ±
and is specified for 1 year after calibration.
Condition of Auto Power On :
With batteries fitted :
- > 3.0V or < -8.0V between L2 and L1
- Detect AC signal by Single Pole
- Continuity
Auto Power Off :
The Meter automatically turns off if one of the following conditions are met
for about 10 seconds
- The Auto Power On condition is not met.
- Both buttons are not pressed.
The Meter automatically turns off if one of the following conditions are met
for about 30 seconds
- The resistance is OL when the Meter is in Resistor mode.
- The current is < 1.0A when the Meter is in Ampere mode
• For > 300V, Time rating (tr): 30 seconds; Recovery time (rt): 240 seconds
• AC Function
- ACV and ACA specifications are ac coupled, true RMS.
- For non-sinusoidal waveforms, Additional Accuracy by Crest Factor (C.F.) :
Add 1.0% for C.F. 1.0 ~ 2.0
Add 2.5% for C.F. 2.0 ~ 2.5
Add 4.0% for C.F. 2.5 ~ 3.0
- Max. Crest Factor of Input Signal:
3.0 @ 5000 counts
1.5 @ 10000 counts
• DC Voltage
Range
With batteries
7.0V to 999.9V
(1)
Without batteries
35V to 999.9V
(1) Measurement without batteries is only available for < 35°C,
Max. Input Current : < 3.5mA @ 1000V
Overload Protection : AC/DC 1000V
• AC Voltage
Range
With batteries
6.0V(1) to 999.9V
(2)
Without batteries
45V to 999.9V
(2) Measurement without batteries is only available for < 35°C,
Frequency Response :
Max. Input Current : < 3.5mA @ 1000V
Overload Protection : AC/DC 1000V
EN
Without batteries :
> | ±35.0V DC | or > 45.0V AC
between L2 and L1
Resolution
Accuracy
0.1V
±(1.0% + 2D)
0.1V
Resolution
Accuracy
0.1V
±(1.5% + 5D)
0.1V
07