Télécharger Imprimer la page

EGAmaster EGAtronik 51252 Manuel D'instructions page 23

Publicité

Les langues disponibles

Les langues disponibles

Range
200.0µA
2000µA
20.00mA
200.0mA
2.000A
10.00A
-
Frequency response: 40-400Hz.
-
Accuracy guarantee range: 5 -100% of the range, shorted circuit allows least significant
digit
2.
-
Input curren! >10.10A, "OL" symbol appears with beeps.
-
Overload protection 250Vrms
µA mA range : F1 Fuse 0.2A/250V
10A range: F2 Fuse 10A/250V
Warning: Before opening the rear cover, switch off the power supply (remove test leads
from the input terminal and the circuit).
GENERAL MAINTENANCE
1. Clean the case with a damp doth and detergent. Do not use abradants or solvents.
2. lf there is any malfunction, stop using the device and send it to maintenance.
3. The maintenance and service must be conducted by qualified professionals or
designated departments.
BATTERY REPLACEMENT:
To avoid false reading, replace the battery when the battefy indicator
Specification: AAA 1.5Vx2.
1. Switch the dial to "OFF "positionand remove the test leads from the input terminal.
2. Take off the protective case. Loosen the screw on battery cover, remove the cover to
replace the battery. Please identify the positive and negative pole.
FUSE REPLACEMENT:
1. Switch the dial to "OFF "position and remove the test leads from the input terminal.
2. Loosen the both screws on the rear cover, then remove the rear cover to replace the
fuse.
AC CURRENT
Resolution
0.1µA
1µA
0.01mA
0.1mA
0.001A
0.01A
Φ
5x20 mm
Φ
5x20mm
MAINTENANCE
REPLACEMENTS (see Figure 4)
Accuracy
±
(1.2%+3)
±
(1.2%+3)
± (1.2%+3)
±
(1.2%+3)
±
(1.5%+5)
± (1.5%+5)
appears. Battery
23

Publicité

loading