Miscellaneous; Error Messages - Medisana CardioDock Mode D'emploi

Module de mesure de tension
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 76

6 Miscellaneous

6.1 Error messages

Unusual reading
- Repeat the test.
- The cuff is not positioned correctly or the cuff
is too small.
-
You did not keep to the correct posture during
the reading
-
During the reading, you spoke, moved your arm
or body or you are excited or nervous.
-
If you get another unusual reading, contact your
doctor.
Irregular pulse (arrhythmia)
-
This device is not suitable for persons with an
acute irregular pulse.
The cuff does not pump up properly
- Check that the air hose is fitted correctly.
- The air hose is pinched or is leaking.
- The cuff is leaking.
The device does not respond
-
The Apple device is not connected properly to
the blood pressure module.
- The battery is too weak.
- Separate the Apple device and the module, re-
start the software and reconnect the Apple device
to the module.
- The LED of the start/stop-button flashes alternately
yellow and blue: The CardioDock battery is defective.
30
GB
.

Publicité

Table des Matières
loading

Table des Matières