Recommended separation distances between RF wireless communications equipment
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between
RF wireless communications equipment and the device as recommended below, according to the maximum output power of the
communications equipment.
Frequency MHz
Maximum
Power
W
385
1.8
450
2
710
745
0.2
780
810
870
2
930
1720
1845
2
1970
2450
2
5240
5500
0.2
5785
Note 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
WARNINGS
•
This device should not be used in the vicinity or on the top of other electronic equipment such
as cell phone, transceiver or radio control products. If you have to do so, the device should be
observed to verify normal operation.
•
The use of accessories and power cord other than those specified, with the exception of
cables sold by the manufacturer of the equipment or system as replacement parts for internal
components, may result in increased emissions or decreased immunity of the equipment or
system.
Distance
IEC 60601
Test Level
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9
Compliance
Electromagnetic Environment - Guidance
Level
27
RF wireless communications equipment
should be used no closer to any part
28
of the device, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
9
Recommended separation distance
28
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
28
recommended separation distance in meters
(m). Field strengths from fixed RF transmitter,
as determined by an electromagnetic site
survey, should be less than the compliance
28
level in each frequency range. Interference
may occur in the vicinity of equipment marked
with the following symbol:
9
125