Les langues disponibles

EN-10
3.3 Temperature Calibration (Cont.)
5.
START CALIBRATION
160C
TARE
6.
AFTER 15 MINUTE TIME PERIOD
7.
ACCEPT CALIBRATION
8.
FINISH CALIBRATION
TARE
3.4 Time and Date Setting
1.
TO START
2.
SET FORMAT
TARE
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
CURRENT TEMP:
100 C
TM TO CAL PNT:
15:00 MIN
ADJ CAL READING: 160C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
CURRENT TEMP:
160 C
TM TO CAL PNT:
10:00 MIN
ADJ CAL
READING:
161 C
ACCEPT NEW CAL
ABORT CALIBRATION
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
or
15 minute time lapse
or
ADJUST TO THERMOMETER READING
TARE
Open the cover, allow unit to cool, remove the
calibration fixture, replace the pan support and pan.
x2
TIME - DATE
TIME - DATE
TIME - DATE
TIME - DATE
TIME - DATE
FORMAT:
MM/DD/YR
SET DATE:
07/10/01
TIME FMT:
12HR
SET TIME:
11:30 AM
MB35
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
CURRENT TEMP:
160 C
TM TO CAL PNT:
10:00 MIN
ADJ CAL READING:
160C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
TEMP CAL 160C
CURRENT TEMP:
160 C
TM TO CAL PNT:
10:00 MIN
ADJ CAL READING:
161 C
ACCEPT NEW CAL
ABORT CALIBRATION
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TIME - DATE
TIME - DATE
TIME - DATE
TIME - DATE
TIME - DATE
FORMAT:
DD.MM.YR
SET DATE:
10. 07. 01
TIME FMT:
12HR
SET TIME:
11:30 AM