Failure
LED of the on/off switch
remains off when
switching on
Shaking movement stops
In case of recurring errors please contact the responsible sales
department or our technical service. Contact address see "Contact
information Heidolph international" on page 54.
Cause
No mains voltage
LED defective
Timer expired
Motor overheat protection
has tripped
Electrical fault (no engine
noise)
Mechanical defect (engine
noise audible)
47
Possible remedy
Check the power supply cord
for correct connection
Check RCD
Check fuses of the device
Contact Heidolph Sales or
Service
Restart the timer if necessary
Wait approx. 20 minutes
and reduce the load on the
shaking platform
Contact Heidolph Sales or
Service
Contact Heidolph Sales or
Service