Operating Messages; Error Codes - Lovibond MD 200 Mode D'emploi

Photometer-system
Masquer les pouces Voir aussi pour MD 200:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Operating messages – Error codes
GB
Lo
btLo
Store
Date
Cal
RESULT
Time
Cal
E27 / E28 / E29
E 10 / E 11
E 20 / E 21
E23 / E24 / E25
E 22
E 70
E 71
E 72
E 73
E 74
E 75
MD200_3 02/2014

Operating messages

Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.
A user calibrated method is indicated by an arrow while the test result
is displayed (see "Factory calibration reset").

Error codes

Light absorption too great. Reasons: e.g. dirty optics.
Calibration factor "out of range"
Too much light reaching the detector.
Too much light reaching the detector.
Battery capacity was too low during measurement. Change battery.
HP1:
Factory calibration incorrect / erased
HP1:
User calibration incorrect / erased
HP2:
Factory calibration incorrect / erased
HP2:
User calibration incorrect / erased
pH:
Factory calibration incorrect / erased
pH:
User calibration incorrect / erased
39

Publicité

Table des Matières
loading

Table des Matières