Functional Description; Operation; Otz (One Time Zero) - Tintometer Lovibond MD200 Mode D'emploi

Masquer les pouces Voir aussi pour Lovibond MD200:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Functional description

GB

Operation

On
Switch the unit on using the [ON/OFF] key.
Off
METHOD
The display shows the following:
Select the required test using the [MODE] key.
Mode
Scroll Memory (SM)
To avoid unnecessary scrolling for the required test method, the instru-
ment memorizes the last method used before being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test method first.
METHOD
The display shows the following:
Fill a clean vial with the water sample up to the 10 ml mark, screw
the cap on and place the vial in the sample chamber making sure that
the
Zero
Press the [ZERO/TEST] key (see OTZ).
Test
The "Method" symbol flashes for approx. 8 seconds.
METHOD
The display shows the following:
0.0.0
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the
Zero
Press the [ZERO/TEST] key.
Test
The "Method" symbol flashes for approx. 3 seconds.
METHOD
RESULT
The result appears in the display.
The result is saved automatically.
Repeating the test:
Zero
Test
Press the [ZERO/TEST] key again.

OTZ (One Time Zero):

The zero setting is held in memory until the instrument is switched
off. It is not necessary to perform a new zero each time, if the water
samples under test are from the same body of water and the conditions
of testing are the same.
The zero setting can be repeated each time if necessary.
Repeating the zero:
Zero
Test
Press the
22
All manuals and user guides at all-guides.com
marks are aligned.
marks are aligned.
[ZERO/TEST]
key for 2 seconds.
MD200_1 09/2011

Publicité

Table des Matières
loading

Table des Matières