Correct Use; Unpacking; Commissioning; Setting The Target Temperature - IKA HB digital 092017 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

• Use
The IKA tempering baths HB digital is laboratory device and is
suitable for directly tempering substances filled into the bath con-
tainer.
They are also suitable for indirect tempering of substances filled
in glass containers when the glass containers are immersed in the
actual tempering medium. A rotary glass container is particularly
advantageous, for example, when used in conjunction with an
IKA rotary evaporator.
The device is not intended to pre-
CAUTION
pare food!

• Unpacking

- Unpack the device carefully,
- Any damage should be notified immediately to the shipping
agent (post, rail or logistics company).
Observe the ambient conditions (temperature, humidity, etc.) list-
ed under "Technical Data".
Prior to commissioning the device, determine the filling quantity
of the tempering medium due to the used evaporator piston size
(approx. 2.5 l tempering medium by using a standard 1 l evapora-
tor piston).
• Switch on the device using the mains switch (B).
• At every start, the display will show all display segments (Fig. 2)
and the software version (Fig. 4).

Setting the target temperature

Activate the SET function by turning and pressing the rotating/
pressing knob (C).
• Switch on the device using the mains switch (B), when the ope-
ning screen (Fig. 5) appears and the setting for the safety tem-
perature is activated for 5 seconds. "SAFE TEMP." flashes on the
display, see also "Setting the safety temperature".
• The display then switches from "SAFE TEMP." to "TEMP." and the
SET function is activated. (Fig. 6)

Correct use

Unpacking

Commissioning

The unit is ready for service when
the mains plug has been plugged
in.
• Area of use (only indoors)
- Laboratories
- Pharmacies
This device is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply net-
work that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not supplied or
recommended by the manufacturer;
- If the device is operated improperly or contrary to the manufac-
turer's specifications;
- If the device or the printed circuit board are modified by the third
parties.
• Scope of delivery
- Heating bath IKA HB digital
- Power supply cable
- USB cable
- Operating instructions
- Warranty card
Note: Only the display segments shown on Fig. 3 are available
in functions for the current device.
• The set operating mode (A/B/C) is indicated on the top of the
display (Fig. 5).
• The heating bath temperature is held constant by the control
circuit and is additionally monitored by the safety circuit.
In the event of an error occurred in the control circuit, the heat-
ing bath is switched off permanently by the safety circuit.
An error in the control or safety circuits is shown on the display.
The heating function can no longer be started.
• When switching off the heating function and the medium tem-
perature is higher than 50 °C, the display will alternately show
OFF and HOT (Fig. 6, Fig. 8).
• Set the desired heating bath temperature by turning the rota-
ting/pressing knob (C).
• The SET function flashes every few seconds.
• Confirm the value by pressing the rotating/pressing knob (C).
• If not confirmed, the set value is not saved.
• The SET function stops flashing.
• Start the heating process by pressing the rotating/pressing knob
(C).
- Schools
- Universities
6

Publicité

Table des Matières
loading

Table des Matières