5 MAINTENANCE
5.7
Troubleshooting
FAULT
No motor sound audible/the lift
does not move
Lift can only be raised
Lift stops when rising
LED on power socket does not
light up during charging
Device slips
LED on hand control unit lights up
during operation
Invacare® Rio™ H605
POSSIBLE CAUSE
Battery is empty or not sufficiently charged
Battery contacts are dirty
Battery is not inserted correctly
Cable for hand control unit is damaged.
Battery not sufficiently charged
Maximum load exceeded
Battery defective
Charger not inserted correctly
Charger defective
Battery is already fully charged
Suction cups damaged
Battery not sufficiently charged
Electronics defective
Recharge battery fully
Clean battery contacts
Insert battery correctly
Contact a qualified technician.
Recharge battery fully
Observe maximum load
Replace battery
Insert charger correctly
Replace charger
-----
Replace suction cups
Recharge battery fully
Contact a qualified technician.
16
REMEDY