Important Notes; Recommandations Importantes; Service De Calibrage - Ebro TFI 550 Mode D'emploi

Table des Matières

Publicité

English
3.14 Important Reminders
Not recommended for use in measu-
ring shiny or polished metal surfaces
(stainless steel, aluminum, etc.). See
"Emissivity."
The unit cannot measure through
transparent surfaces such as glass or
plastic. It will measure the surface tem-
perature of the glass instead.
Steam, dust, smoke, etc., can pre-
vent accurate measurement by obstruc-
ting the unit's optics.
When measuring liquid products, be
sure to stir the product vigorously while
measuring the surface of the liquid.
Calibration service
4
To guarantee a high accuracy of measu-
rements, the thermometer TFI 550 must
be calibrated once annually. ebro Elec-
tronic GmbH & Co. KG offers a service
for this.
Français
3.14 Rappels
Il est déconseillé d'utiliser ce ther-
momètre pour mesurer la température
de surfaces métalliques brillantes ou
polies (acier inoxydable, aluminium,etc.).
Voir Emissivité.
Le thermomètre ne peut pas mesurer
la température à travers des surfaces
transparentes comme le verre, car il
mesure en fait la température de surface
du verre.
La précision des résultats peut être
faussée par la présence de vapeur, de
poussière, de fumée, etc.
Lorsque vous mesurez des liquides,
veillez à les brasser vigoureusement
avant d'en mesurer la surface.
4

Service de calibrage

Pour garantir une bonne précision de la
mesure, le thermomètre à infrarouge TFI
550 doit être calibré une fois par an. Pour
cela, ebro Electronic GmbH & Co. KG
vous propose ses services.
25

Publicité

Table des Matières
loading