1. Attach the electrodes to the cables.
If carbon rubber electrodes are being
used, spread on an even layer of gel.
Wash and dry the skin well before
attaching the electrodes.
2. Turn the stimulator on with
you wish to use the same settings as
last time you stimulated, you can
start stimulation directly by setting
the desired amplitude with
each channel.
3. To choose programs press
until the wanted program is shown.
Program:
Stimulation form:
P1
Conventional TENS
P2
Burst TENS
P3
Modulated pulse duration TENS

REPLACEMENT OF BATTERIES

When? A battery indicator is lit on the
display when the batteries are almost
empty. As long as stimulation feels the
same as usual you can continue the
treatment. It is time to replace the
batteries when stimulation starts
diminishing or the stimulator turns off.
NOTE! Non-rechargeable batteries may
not be charged due to the risk of
explosion.
NOTE! This can only be done when
both channels are set to 0.0 mA.
4. Start stimulation by selecting the
desired amplitude with
channel.
. If
NOTE!
20 seconds after the amplitude has been
set the key lock is automatically
for
activated. Deactivate the key lock by
pressing one of the
Frequency:
80 Hz
2 Hz
80 Hz
How? Turn off the stimulator. Remove
the battery cover by sliding it downwards
and then lifting it upwards and take out
the batteries. Make sure the new batteries
are placed correctly with respect to the
polarity (note marks for + and –). See p.
97.
NOTE! Do not throw the batteries out,
but recycle appropriately.
for each
buttons.
Pulse duration:
180 µs
180 µs
70–180 µs
21