5 SCPI Programming Reference
Calibration mode is locked out by the calibration switch.
104 Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.
105 Unexpected output current
The measured output current is outside the acceptable range.
106 Zero measurement out of range error
The "zero" measurement value is outside the acceptable range.
107 Programming cal constants out of range
The programmed calibration constant is outside the acceptable range.
108 Measurement cal constants out of range
The measurement calibration constant is outside the acceptable range.
109 Over voltage cal constants out of range
The over voltage calibration constant is outside the acceptable range.
110 Wrong V+I
The instrument was unable to set the correct voltage or current value.
114 Wrong status
An incorrect status function has been reported.
116 Locked out by internal switch setting
This function has been locked out by an internal switch.
117 Calibration error
A calibration error has occurred. Do not save calibration constants. Try re-calibrating the unit.
200 Hardware error channel <1>
A hardware error has occurred on the output.
201 Invalid configuration
An invalid parallel or SDS configuration is not allowed.
202 Selftest Fail
A selftest failure has occurred. See selftest failure list for details.
203 Compatibility function not implemented
The requested compatibility function is not available.
204 NVRAM checksum error
Keysight RP7900 Series Operating and Service Guide
265