Keysight Technologies RP7900 Serie Guide D'utilisation Et De Maintenance page 210

Masquer les pouces Voir aussi pour RP7900 Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5 SCPI Programming Reference
Parameter
(none)
Return a 1 when commands complete: *OPC?
The purpose of this command is to synchronize your application with the instrument.
l
Other commands cannot be executed until this command completes.
l
*OPT?
Returns a string identifying any installed options. A 0 (zero) indicates no options are installed.
Parameter
(none)
Returns installed options *OPT?
*RCL <0-9>
Recalls a saved instrument state. This restores the instrument to a state that was previously stored in
locations 0 through 9 with the *SAV command. All instrument states are recalled except: (1) the output
state is set to OFF, (2) the trigger systems are set to the Idle state, (3) calibration is disabled, (4) all
lists are set to their *RST values, and (5) the non-volatile settings are not affected.
Parameter
0 - 9
Recall state from location 1: *RCL 1
Location 0 is automatically recalled at power turn-on when the Output Power-On state is set to
l
RCL 0.
Stored instrument states are not affected by *RST.
l
*RST
Resets the instrument to pre-defined values that are either typical or safe. These settings are
described in
Reset
State.
Parameter
(none)
Reset the instrument: *RST
*RST forces the ABORt commands. This cancels any measurement or transient actions presently
l
in process. It resets the WTG-meas, MEAS-active, WTG-tran, and TRAN-active bits in the Oper-
ation Status registers.
210
Typical Return
1
Typical Return
OPT 760
Typical Return
(none)
Typical Return
(none)
Keysight RP7900 Series Operating and Service Guide

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Rp7930aRp7940aRp7950aRp7960aSd1000a

Table des Matières