OUTPut:INHibit:MODE LATChing|LIVE|OFF
OUTPut:INHibit:MODE?
Sets the operating mode of the remote inhibit digital pin. The inhibit function shuts down the output in
response to an external signal on the Inhibit input pin. The Inhibit mode is stored in non-volatile
memory. See
Programming the Digital
LATChing - a logic-true signal on the Inhibit input causes the output state to latch OFF. The output
remains disabled until the Inhibit input is returned to logic-false and the latched INH status bit is
cleared by sending the OUTPut:PROTection:CLEar command or a protection clear command from the
front panel.
LIVE - allows the enabled output to follow the state of the Inhibit input. When the Inhibit input is true,
the output is disabled. When the Inhibit input is false, the output is re-enabled.
OFF - The Inhibit input is ignored.
Parameter
LATChing|LIVE|OFF
Sets the Inhibit Input to Live mode: OUTP:INH:MODE LIVE
OUTPut:PON:STATe RST|RCL0
OUTPut:PON:STATe?
Sets the output power-on state. This determines whether the power-on state is set to the *RST state
(RST) or the state stored in memory location 0 (RCL0).Instrument states can be stored using the *SAV
command. This parameter is saved in non-volatile memory.
Parameter
RST|RCL0
Sets the power-on state to the *RST state: OUTP:PON:STAT RST
If the power-on state is set to 0 with no state stored, a self-test error "file not found; 0 state" is gen-
l
erated and the instrument is set to the *RST state.
If a primary
auto-connect
l
OUTPut:PROTection:CLEar
Resets the latched protection. This clears the latched protection status that disables the output when
a protection condition occurs (see
Parameter
(none)
Clears the latched protection status: OUTP:PROT:CLE
Keysight RP7900 Series Operating and Service Guide
Port.
Typical Return
LATC, LIVE, or OFF
Typical Return
RST or RCL0
command fails, the power-on state is set to *RST.
Programming Output
Typical Return
(none)
5 SCPI Programming Reference
Protection).
225