Functional description

GB
On
Off
METHOD
Mode
METHOD
Zero
Test
METHOD
0.0.0
Zero
Test
METHOD
RESULT
Zero
Test
Zero
Test
20

Operation

Switch the unit on using the [ON/OFF] key.
The display shows the following:
Select the required test using the [MODE] key.
The display shows the following:
Fill a clean vial with the water sample up to the 10 ml mark, screw
the cap on and place the vial in the sample chamber making sure that
the
marks are aligned.
Press the [ZERO/TEST] key.
The "Method" symbol flashes for approx. 8 seconds.
The display shows the following:
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the
marks are aligned.
Press the [ZERO/TEST] key.
The "Method" symbol flashes for approx. 3 seconds.
The result appears in the display.
The result is saved automatically.
Repeating the test:
Press the [ZERO/TEST] key again.
Repeating the zero:
Press the
[ZERO/TEST]
Caution:
A calibration of the method has to be performed before the first use
(see
"menu-options"
of SPADNS reagent solution must be used for adjustment and test.
The adjustment process needs to be performed for each new batch of
SPADNS reagent solution.
key for 2 seconds.
4
calibration mode Fluoride). The same batch
MD100_5 10/2014