● Should be as small as possible (short lead length) in order to maintain high
signal integrity.
Fig. 5-4: Ground connection and connection inductance using the example of R&S RT-
ZD10/20/30
The
table 5-2
shows different types of connections between the probe and DUT
as well as the associated connection inductance L
ical bandwidth values, rise time values, and the step responses for each connec-