RT-ZD40
Bandwidth
Dynamic range
(differential input)
Operating voltage window
(each pin to GND)
Diff. input resistance
Diff. input capacitance
R&S ProbeMeter, measurement error
Extremely low zero and gain errors throughout the entire temperature range, no significant tem-
perature drift
Micro button
Rohde & Schwarz probe interface
1.1.2

Key Features

Micro button
The micro button is located at the probe head. The micro button can be config-
ured via the base unit. By pressing it, the user can remotely control different func-
tions on the base unit. For details, see
R&S ProbeMeter
The R&S ProbeMeter measures the DC voltage of the input signal directly at the
probe tip. It provides a continuous high-precision DC voltage measurement that is
independent of the settings of the oscilloscope and in parallel to the time domain
measurement. If activated on the base unit, the measurement value is displayed
on the screen of the R&S oscilloscope. For details, see
Meter",
on page 21.
The R&S ProbeMeter records both differential and common mode DC voltages.
Data memory
The probe includes an integrated data memory with individual probe correction
parameters (e.g. gain, delay, offset). These correction parameters are read out
and processed by the R&S oscilloscope. As a result, the probe offers a very high
degree of accuracy so that additional calibration procedures are not required.
User Manual 1410.5363.02 ─ 03
Key Features and Key Characteristics
DC to 4.5 GHz
±5 V with ±5 V offset capability
10 V AC (Vpp)
±8 V with ±22 V common mode offset capability
available for R&S RT-ZD40 probes with serial number
≥ 200000
1 MΩ
0.4 pF
<0.1 %
chapter 2.2, "Micro
Product Description
Button",
on page 18.
chapter 2.4, "R&S Probe-
6